Home




Ozone Gas

Sensor Head Specifications

OZONE gas sensor heads are available in a number of application focused configurations namely:

  • Low Concentration Sensor Heads
  • High Concentration Sensor Heads
  • Leak Detection Sensor Heads
  • Controlled Atmosphere Sensor Heads
  • Custom Sensor Heads

Aeroqual also offers the above sensor heads specially calibrated for "Controlled Atmosphere" applications, typically for low oxygen environments. In addition customised calibration to customer requirements is available on request.

Sensor Heads

Low

High

Leak

Measurement range

0.000 - 0.500 ppm

0.50 - 20.00 ppm

0.00 - 50.00 ppm

Accuracy

< 0.008 ppm (0 - 0.100 ppm)

± 10% (0.100 – 0.500 ppm)

± 10% (0.20 - 2.00 ppm)

± 15% (2.00 - 20.00 ppm)

± 20%,

15 seconds after Reset

Resolution

0.001 ppm

0.01 ppm

0.01 ppm

Controlled Atmosphere Version

Available on Request

Available on Request

Available on Request

T90 Response

< 60 seconds (T90)

< 35 seconds (T90)

< 10 seconds

Sensor type

Gas-sensitive semiconductor

Operating temperature range

-5°C to 50°C; 23°F to 122°F (sensor head and base unit)

Relative humidity limit

95% maximum (sensor head and base unit)


Ozone Gas
Monitoring Products

OZONE gas sensor heads are compatible with all Aeroqual monitoring products.

The sensor heads conduct "Active Air Sampling" of ambient air and utilize the proprietary Aeroqual "Gas Sensitive Semiconductor" (GSS) technology.

All Aeroqual products have removable sensor heads that eliminate the need for field calibration and facilitate sensor head replacement as required.

Select one of the Product Categories from the links on the right to access the appropriate product selection chart.



 
Hand-held, Portable Monitoring ProductsSee
Fixed Installation, Monitoring ProductsSee
Network System, Monitoring ProductsSee