SM50 Sensor Module Specifications - Standard Calibration
For Industrial and Health & Safety applications
Sensor |
Calibrated Range (ppm) |
0-5V out scale |
LDL |
Accuracy |
Resolution |
Response Time (T90) |
Sampling Method
(5) |
Operating Temp.
(environ-mental) |
Relative Humidity
(non-condensing) |
Ammonia
0 - 100 ppm |
0 - 100 |
0 - 100 |
0.5 ppm |
± 5ppm 0 - 100 ppm |
0.1 ppm |
< 60 s |
D |
-5°C to 40°C |
5 to 95% |
Ammonia (leak)
0 - 1000 ppm |
0 - 1000 |
0 - 1000 |
2 ppm |
± 15% |
1 ppm |
< 60 s |
D |
-5°C to 40°C |
5 to 95% |
Carbon monoxide
0 - 1000 ppm |
0 - 1000 |
0 - 1000 |
1 ppm |
± 20% |
1 ppm |
< 150 s |
D |
-5°C to 40°C |
5 to 95% |
Carbon dioxide
0 - 2000 ppm |
0 - 2000 |
0 - 5000 |
- |
± (40 ppm + 3%) |
10 ppm |
< 60 s |
D |
-5 to 40°C |
5 to 95% |
Carbon dioxide
0 - 5000 ppm |
0 - 5000 |
0 - 5000 |
- |
± (150 ppm + 5%) |
10 ppm |
< 60 s |
D |
-5 to 40°C |
5 to 95% |
Carbon dioxide
0 - 5.00% |
0 - 5.00% |
0 - 5% |
- |
± 5% |
0.01% |
< 60 s |
D |
-5 to 40°C |
5 to 95% |
Hydrogen
0 - 5000 ppm 1 |
0 - 5000 |
0 - 5000 |
5 ppm |
± 10 % |
1 ppm |
< 90 s |
D |
-5°C to 40°C |
5 to 95% |
Hydrogen sulphide
0 - 50 ppm |
0 - 50 |
0 - 50 |
0.05 ppm |
<+/-10% |
0.1 ppm |
< 60 s |
D |
-5°C to 40°C |
5 to 95% |
Methane
0 - 10000 ppm |
0 - 9999 |
0 - 10000 |
- |
± 15% |
1 ppm |
< 60 s |
D |
-5°C to 40°C |
30 to 80% |
Ozone
0 - 0.5 ppm |
0 - 0.500 |
0 - 0.500 |
1 ppb |
± 15% |
0.001 ppm |
< 60 s |
F,P |
-5°C to 40°C |
5 to 95% |
Ozone
0.5 - 20 ppm 2 |
0.5 - 20 |
0 - 20 |
10 ppb |
± 15% |
0.01 ppm |
< 35 s |
F,P |
-5°C to 40°C |
5 to 95% |
Nitrogen dioxide
0 - 0.2 ppm |
0 - 0.200 |
0 - 0.5 |
1 ppb |
± 15% |
0.001 ppm |
< 60 s |
F,P |
-5°C to 40°C |
30 to 70% |
Perchloroethylene
0 - 200 ppm |
0 - 200 |
0 - 500 |
1 ppm |
± 5 ppm 0 - 50 ppm
± 10% 50 - 200 ppm |
1 ppm |
< 60 s |
D |
-5°C to 40°C |
30 to 80% |
PID
0 - 2000 ppm |
0 - 2000 |
0 - 5000 |
0.1 ppm |
<10% of reading |
0.1 ppm |
<30 s |
D |
-5°C to 40°C |
5 to 90% |
Sulphur dioxide
0 - 100 ppm |
0 - 100 |
0 - 100 |
0.5 ppm |
± 10% |
0.1 ppm |
< 60 s |
D |
-5°C to 40°C |
5 to 95% |
VOC
0-500 ppm |
0 - 500 |
0 - 500 |
1 ppm |
± 10 ppm 0 - 200 ppm
± 10% 200 - 500 ppm |
1 ppm |
< 60 s |
F |
-5°C to 40°C |
5 to 95% |
Other Gases |
Contact us with specific requirements for gas and concentration |
- Other specific concentrations available on request
- Accuracy < 0.5 ppm is unspecified (for applications that require accurate measurement < 0.5 ppm use low or ultra-low sensor heads)
- Non Methane Hydrocarbon sensor (lower sensitivity to alcohols & esters)
- Calibrated against Isobutylene. Other specific VOC calibrations are available on request
- D= diffusion, F = fan, P= pump
top
SM50 Sensor Module Specifications - High Specification Calibration
For ambient measurements
Sensor |
Calibrated Range (ppm) |
0-5V out scale |
LDL |
Accuracy |
Resolution |
Response Time (T90) |
Sampling Method
(5) |
Operating Temp.
(environ-
mental) |
Relative Humidity
(non-condensing) |
Carbon monoxide
0 - 100 ppm |
0 - 100 |
0 - 100 |
0.5 ppm |
± 5 ppm |
0.1 ppm |
< 150 s |
D |
-5°C to 40°C |
5 to 95% |
Hydrogen sulphide
0 - 10 ppm |
0 - 10 |
0 - 10 |
10 ppb |
± 0.5 ppm |
0.01 ppm |
< 60 s |
D |
-5°C to 40°C |
5 to 95% |
Ozone
0 - 0.150 ppm |
0 - 0.150 |
0 - 0.500 |
1 ppb |
± 0.005 ppm |
0.001 ppm |
< 70 s |
F, P1, P2 |
-5°C to 40°C |
5 to 95% |
Ozone
0 - 0.5 ppm |
0 - 0.500 |
0 - 0.500 |
1 ppb |
±0.008ppm 0 - 0.1ppm
± 10% 0.1 - 0.5ppm |
0.001 ppm |
< 60 s |
F, P1,P2 |
-5°C to 40°C |
5 to 95% |
NMHC 1
0 - 25 ppm |
0 - 25 |
0 - 25 |
0.1 ppm |
± 10% 0.1 - 25 ppm |
0.1 ppm |
< 60 s |
D |
-5°C to 40°C |
5 to 95% |
PID
0 - 20 ppm |
0 - 20 |
0 - 50 |
0.01 ppm |
<10% |
0.01 ppm |
<30 s |
D |
-5°C to 40°C |
5 to 90% |
Sulphur dioxide
0 - 10 ppm |
0 - 10 |
0 - 10 |
0.2 ppm |
± 0.5 ppm |
0.01 ppm |
< 60 s |
D |
-5°C to 40°C |
5 to 95% |
VOC
0 - 25 ppm |
0 - 25 |
0 - 25 |
0.1 ppm |
± 10% 0.1 - 25 ppm |
0.1 ppm |
< 60 s |
D |
-5°C to 40°C |
5 to 95% |
Temperature |
-20 to 100°C |
-40 to 120°C |
0.01°C |
± 0.3°C |
0.01°C |
< 1 s |
N/A |
-40 to 120°C |
0 to 100% |
Humidity |
0 to 100% |
100% |
1% RH |
2% RH |
1% RH |
< 1 s |
N/A |
-40 to 120°C |
0 to 100% |
- Non Methane Hydrocarbon sensor (lower sensitivity to alcohols & esters)
- D=diffusion, F=fan, P1=economy pump, P2 = longlife pump
Note: These units come with 12 bit DAC on analog outputs as standard. top
|