IQM 60 Indoor Air Quality MonitorIQM 60 Indoor Air Quality Monitor

The complete instrument for analysis of indoor air quality

The IQM 60 Indoor Air Quality Monitor by Aeroqual enables simultaneous monitoring of common air quality parameters with a single instrument.

IQM 60 Indoor Air Quality MonitorAeroqual’s proprietary Analytic Gas Sensitive Semiconductor (GSS) Technology plus photo-ionization detector (PID) and non-dispersive infra-red (NDIR) optical sensors are used to achieve cost effective and precise measurement of Carbon Dioxide (CO2), Carbon Monoxide (CO), Total Volatile Organic Compounds (TVOC), Nitrogen Dioxide (NO2), Ozone (O3), Temperature, Humidity and Particulate Matter.

Air quality data is logged to a removable SD card or PC and can be exported to other programs for generating reports. Applications include IAQ and HVAC analysis, IAQ complaint investigation, and Sick Building Syndrome (SBS) assessment. The IQM 60 offers the complete package for performing indepth analysis of indoor air quality.

Features

  • Multiple Gas Measurement Sensors
  • Analytic GSS Technology
  • Optical Sensor Technology (PID / NDIR)
  • Factory Certified - NIST Traceable
  • Temperature and Humidity Sensors
  • Particulate Measurement Options
    - Particulate Profiler (Dual or 8-channel counter)
    - Particle Monitor (TSP, PM1, PM2.5 or PM10)
  • Active-air Sampling System
  • Zero and Span Calibration Facility
  • Rapid Real-time Data Sampling
  • Large Removable Data Storage
  • PC Data-logging Software (included)
  • Wireless Communication (optional)
  • Compact and Portable
  • Low Maintenance

Applications

  • IAQ Complaint Investigation and Analysis
  • HVAC System Performance Monitoring
  • Sick Building Syndrome (SBS) Surveys
  • Health and Comfort Assessment
  • Odour Investigation and Remediation
  • Testing the Efficiency of Air Purifiers
  • Clean Room Monitoring, Verification and Filter Testing
  • Process Control Monitoring
  • Residential and Commercial Buildings
  • Airport Lounges and Shopping Malls
  • Schools and Kindergartens
  • Hospitals and Elderly Care Facilities
  • Epidemiological Studies
  • Re-entrainment Studies

top

Specifications

Environment Operating Range

0°C to 50°C
5 to 95% RH (non condensing)

Sampling Method

Diaphragm pump and built-in zero-air scrubber with replaceable media

Inlet Filter Replaceable PTFE 5 µm

Zero / Span Calibration

PC configuration software supplied

Communication

RS 232 | Cable and USB adaptor supplied

Data Storage

Removable SD card 1GB

Data Sampling Rate

Programmable frequency 2-255 minutes

Display

4 x 20 vacuum florescent display (VFD)
Scrolling data display

Power Requirements

12VDC / 1A (subject to configuration)
100-250 VAC switch-mode PSU supplied

Enclosure

Aluminium and powder coated steel
278L x 236W x 132H (mm)

Weight

3.6 - 7.5 Kg (subject to configuration)

Communications (optional)

GSM modem | Wireless

External Battery (optional) Li-ion portable power station and charger
(specification subject to configuration)

Accessories

Carry Case
Calibration Gas Humidifier

Maintenance Checks Routine

6-monthly and 12-monthly

top


Gas Sensor Modules

Gas Sensor Modules (code)

Sensor

Range

Lowest Detection

Accuracy

Precision

Resolution

Carbon dioxide (CD) NDIR 0 - 2000 ppm
6 ppm < 40 ppm + 3% 6 ppm 1 ppm
Carbon dioxide (CE) NDIR 0 - 5000 ppm 6 ppm < 150 ppm + 5%
6 ppm 1 ppm
Carbon monoxide (CN) GSS 0 - 100 ppm 0.2 ppm <± 2 ppm <20;
<± 10% >20ppm
1 ppm 0.1 ppm
Total VOC IsoB (VM) GSS 0 - 25 ppm 0.1 ppm <± 10% 0.2 ppm 0.1 ppm
Total VOC IsoB (PD) PID 0 - 20 ppm 10 ppb < 10% 0.02 ppm 0.01 ppm
Non-methane hydrocarbon (VN) GSS 0 - 25 ppm 0.1 ppm <± 0.5 ppm 0.2 ppm 0.1 ppm
Hydrogen sulphide (HS) GSS 0 - 10 ppm 10 ppb <± 0.5 ppm 0.02 ppm 0.01 ppm
Nitrogen dioxide (NW) GSS 0 - 0.200 ppb 1 ppb <± 0.010 ppm 0.005 ppm 0.001 ppm
Ozone (UZ) GSS 0 - 0.150 ppb 1 ppb <± 0.005 ppm 0.002 ppm 0.001 ppm
Ozone (LZ) GSS 0 - 0.500 ppb 1ppb <± 0.008 ppm 0.004 ppm 0.001 ppm

* Contact us for other calibrations available

top

Temperature & Relative Humidity Probe

Sensor Range Accuracy Resolution

Temperature sensor

-20°C to 100°C

± 0.3°C @ 25°C

0.01°C

Relative humidity sensor 0 - 100% RH ± 2% RH @ 25°C 0.1% RH

top

Particulate Profiler Options

Sensor Counter Options Range Accuracy Sensitivity Resolution

Right angle light scatter principle

Dual-channel:
>0.3µm | >2.5µm

0-100,000 particles/L

± 10%

0.3µm

Particles/L

Isokinetic inlet Eight-channel:
0.3µm to 10µm
  to calibration aerosol    

top

Approvals

EN50082-1:1997 | EN50081-1:1992
Part 15 FCC Rules | IPC A 610D Class 2

top