AQM 60 Outdoor Air Quality AQM 60 Outdoor Air Quality MonitorMonitor

The New Benchmark for Cost Effective Air Quality Monitoring

AQM 60 Ambient Air Quality Monitors fill the gap between full reference monitoring stations and simple diffusion samplers. Typical applications include traffic or roadside monitoring, urban air quality measurement and fenceline monitoring of air pollution sources. Aeroqual utilises proprietary Analytic GSS Technology to provide high performance within a flexible cost effective package.

AQM 60 monitors are discreet, highly compact, and deliver precise measurement data for a wide range of ambient pollutants including O3, NO2, CO, SO2, CO2, VOC, plus PM10 and environmental parameters such as temperature, humidity, wind speed and direction, noise and other meteorological sensors. Data is stored internally and is also available in real-time via GSM modem, low power radio, Ethernet and other communication options.

Features

  • AQM 60 Outdoor Air Quality Monitor Analytic GSS Technology
  • Multiple gas sensors (up to six)
  • Temperature and humidity sensors
  • Rapid real-time data sampling (2-minute)
  • Large data storage capacity (>15 years)
  • RS 232 with PC data logging software
  • Zero and span calibration facility
  • Remote diagnostics capabilities
  • Weatherproof and compact enclosures
  • Thermal management system
  • Particulate monitor (PM2.5 or PM10) *
  • Wind speed and direction anemometer *
  • Noise and meteorological sensors *
  • Modem (RF, GSM, GPRS, UMTS) or Ethernet (LAN, WLAN) communication *

* Optional features available

Applications

  • Urban air quality monitoring
  • Roadside and tunnel monitoring
  • Fenceline and point source monitoring
  • Trend analysis and hotspot screening
  • Substitution of manual sampler networks
  • Commercial indoor air quality monitoring

top

Specifications

Environment Operating Range

10°C to 25°C (standard)
-20°C to 45°C (with heater | cooler)
10 to 90% RH (non condensing)

Power Requirements

12VDC (power subject to configuration)
AC power module 100 - 250 VAC

Maintenance Checks Routine

3 | 6 | 12 months (site dependent)

Standard Enclosure

Fibre reinforced polycarbonate; IP66
515H x 415W x 230D (mm)
(size subject to configuration)

Weight

15-25 Kg (subject to configuration)

Communications Options

GSM modem | RF modem | Ethernet LAN

Other Sensor Options

Noise sensors | Particle counters
Meteorological sensors

Enclosure Options Insulated roadside boxes
Mounting accessories
Metal enclosures
Approvals

Electrical Safety and Conformity
CE; UL E215312
Gas Sensors
EN50082-1:1997; EN50081-1:1992
Part 15 FCC Rules; IPCA 610D Class2
Particle Monitor
Class 1 laser; IEC 60825-1:1998
GSM Modem
CE 0681; Group 3 FAX support Class 1/2

top


Gas Modules


Calibrated Range

Lowest Detection

Accuracy

Precision

Resolution

Power
Ozone * 0 - 150 ppb
1 ppb <± 0.005 ppm
0.002 ppm 0.001 ppm 2.9 W
Nitrogen dioxide
0 - 200 ppb 1 ppb <± 0.010 ppm
0.005 ppm 0.001 ppm 12.3 W
Carbon monoxide 0 - 25 ppm 0.1 ppm <± 1 ppm
0.2 ppm 0.1 ppm 2.5 W
Sulphur dioxide 0 - 10 ppm 0.2 ppm <± 0.5 ppm 0.2 ppm
0.01 ppm 2.5 W
Non-methane hydrocarbon 0 - 25 ppm 0.1 ppm <± 0.5 ppm 0.2 ppm 0.1 ppm 2.5 W
Volatile Organic Compounds
0 - 25 ppm 0.1 ppm <± 10% 0.2 ppm 0.1 ppm 2.5 W
VOC (PID)
0 - 20 ppm 10 ppb <± 10% 0.02 ppm 0.01 ppm 1.5 W
Hydrogen sulphide 0 - 10 ppm 10 ppb <± 0.5 ppm 0.03 ppm 0.01 ppm 2.5 W
Carbon dioxide (NDIR) 0 - 2000 ppm 6 ppm < 40 ppm + 3% 6 ppm 1 ppm 1.0 W

* Also available in 0 - 500 ppb

top

Particle Monitor Module

  Range Lowest Detection Accuracy Precision Sharp Cut Points Power

Forward light scatter nephelometer

0 - 2000 µg/m3

1 µg/m3

8% NIOSH 0600

3 µg/m3 PM1 | PM2.5 | PM10 5.2 W

top

Anemometer Module

  Wind Speed Wind Direction Accuracy Resolution WS Resolution WD Power

Sonic or mechanical combination

0-160 Km/h

0-360°

±3% WS | ±3° WD

0.1 m/s 1.0 W

top

Temperature & Humidity Module

Temperature Humidity Accuracy @ 25°C Resolution T° Resolution RH Power

-20°C to +100°C

0-100% RH

±0.3°C | ±2% RH

0.01°C 0.1% RH 0.5 W

top

Control Module

Data Storage Sampling Rate Communication Display Diagnostics Power

1 GB

2-255 min.

RS 232

4 x 20 Yes 5.8 W

top

Gas Treatment Module

Sampling Pump Sampling Train Zero Air Scrubber Zero Calibration Span Calibration Power

Diaphragm

SS | PTFE

Yes

Auto or manual Yes 1.0 W

top

Thermal Management System

Cooler Heater External Cowling Thermal Insulation Temperature Control Power

Thermoelectric

Yes

Yes

Yes Yes 85-200 W

top